Mass spectrometers : puriFlash® MS

Mass spectrometers : puriFlash® MS

Mass spectrometers puriFlash MS chromatography
Mass spectrometers puriFlash MS chromatography

Configure your
puriFlash® MS
puriFlash® MS-HWM
& MS Splitter

Basic configuration

  • Mass-triggered faction collection via normal or reverse-phase Flash or preparative HPLC
  • Confirmation of the mass and purity of the collected fractions
  • Real-time reaction monitoring
  • Compact mass spectrometer that can be positioned underneath a fume hood, direct analyses.
  • Source ESI, APCI or APCI/ASAP (Option)
  • A single analysis for positive and negative ion mode
  • Flow ESI: 10 µL/min to 1 mL/min
  • Flow APCI: 10 µL/min to 2 mL/min
  • Flow ASAP: 10 µL/min to 2 mL/min
  • puriFlash® MS range: 10 to 1200 m/z
  • puriFlash® MS-HMW range: 10 to 2000 m/z
  • Acquisition frequency per second 10,000 m/z
  • Sensitivity (ESI) 10 pg reserpin (FIA – injection of 5 µL to 100 µL/min)
  • S/N: 100/1 (RMS) with SIM (Single Ion Monitoring) of 609.3 m/z
  • Accuracy +/- 0.1 m/z for the entire acquisition range
  • Stability +/- 0.1 m/z for a period of 12h (18°C -24°C)
  • Polarity-change speed: 50 ms
  • Direct injection (FIA) or ASAP source
  • MS Splitter:
    •  Interchim® concept
    •  Combined “Split and Dilution” Technology
    •  Compatible with all column sizes
    •  Even with a high flow, no counterpressure is generated
    •  Dilution after split to adjust the quantity entering the source and to avoid system saturation
  • Mass: connection to puriFlash® via USB
  • MS Splitter: connection to puriFlash® via RS-232 cable
  • Voltage to ground: 100/240 V AC voltage
  • Outlet of choice: UK, Europe, US

Select your options