AVANT® (U)HPLC
Mass spectrometers : puriFlash® MS
AVANT® (U)HPLC
Modular chromatography systems
Components
- Gradient Pump
Autosampler
UV-Vis detector
Column oven
Solvent organizer
Gradient pumps
- • Quaternary gradient pumps:
HPLC pumping system maximum 10,000 psi (689 bar)
UHPLC pumping system maximum 18,850 psi (1,300 bar)
Flow Rate Range: 1 to 2,000 μL/min
Flow Accuracy: Better than ± 1% or ± 10 µL whichever is greater
Flow Precision: 0.075% RSD
Delay Volume: 691 μL
Includes vacuum degasser and solvent waste tray.
- • Binary gradient pumps :
HPLC pumping system maximum 10,000 psi (689 bar)
UHPLC pumping system maximum 18,850 psi (1,300 bar)
Flow Rate Range: 1 to 4,000 μL/min
Flow Accuracy: Better than ± 1% or ± 10 µL whichever is greater
Flow Precision: 0.075% RSD
Delay Volume: 50 μL
Includes automated solvent selection (2 solvents per pump), high pressure mixing, low volume mixer, vacuum degasser and solvent waste tray
Autosamplers
- • HPLC (10,000 psi)
• HPLC (10,000 psi) with cooling
• UHPLC (18,850 psi)
• UHPLC (18,850 psi) with cooling
- – Capacity:
2 x 96 or 384 well microtiter plates
Or 2 x 48 position 2 mL vials
Optional: 2 x 12 position 10 mL vials
- – Injection modes:
Full loop
Partial loop
µL-pick-up
- – Injection volume:
Programmable from 0 – 9999 µL in 1 µL increments
- – Injection precision:
Full loop injection < 0.3% RSD
Partial loop fill < 0.5% RSD
µL-pick-up < 0.1% RSD
- – Sample viscosity:
0.1 to 5 cP
- – Standard syringe volume :
HPLC 500 µL
UHPLC 250 µL
(other options possible)
- – Standard injection loop :
HPLC 100 µL
UHPLC 50 µL
- – Injection cycle time:
Typically <5s
- – Valve switching time :
60 ms
- – Carry-over :
<0.05% with standard wash
<0.01% with additional wash
- – Sample cooling option :
Minimum: 4 °C ± 2 °C
Maximum: ambient – 3 °C
UV/Vis Detector
- UV-DAD diode array detector
Wavelength range: 190 – 800 nm
Light source: Deuterium discharge lamp, halogen lamp
Scanning speed: 20 Hz
Programmable wavelengths: 8 maximum
Noise level: 3 x 10-6 AU
Spectral half-width: 6 nm
Accuracy adjustment: ± 1 nm
Reproducibility: ± 0.5 nm
Column Oven
- • Operating Range 10 °C above ambient to 90 °C – Column Switching Not included
• Operating Range 10 ºC above ambient to 75 ºC – Column Switching 6-way column selection valve 15,000 psi max
• Optional diverter valve
Accuracy ± 1 °C
Repeatability ± 1 °C
Stability ± 1 °C
Column capacity Maximum 6 x 30 cm columns
Software
Mass Express
A user-friendly, intuitive software platform for instrument control and data acquisitionData Express
A full feature data processing package to interpret and present mass spectral and chromatographic information in the clearest form using the fewest possible steps.Quant Express (optional)
Quant Express is an add-on to the Mass Express software suite that provides a complete, detailed quantitation applicationLC express (optional)
LC Express provides a seamless interface with all Agilent chromatography modules. This, along with Clarity® compatibility, provides integration with nearly all available LC systems.
Minimum Computer Specifications
- Operating System: Windows 10 and 11 – 64 bit operating systems.
Processor: Intel i3 or greater Intel CPU or AMD equivalent
Memory: 8 GB
Hard Disk: >200 GB Hard Drive
Display: Minimum screen size of 1366 x 768 pixels
Ports: 4 USB 2.0 (USB Hub is acceptable)
RJ45 port for Ethernet (or additional USB to RJ45 converter
Power supply
- 110 V or 220 V
Dimensions & weight
-
Pump: Dimensions: (WxDxH) 330 x 540 x 110 mm
Weight: 21 kgAutosampler: Dimensions (WxDxH) 330 x 510 x 360 mm
330 x 575 x 360 mm with cooling
Weight: 19 kg – 21 kg with coolingUV-Vis detector: Dimensions (WxDxH) 330 x 510 x 150 mm
Weight: 9 kgColumn oven: Dimensions (WxDxH) 170 x 600 x 345 mm
Weight: 16 kgSolvent organizer: Dimensions (WxDxH) 330 x 510 x 80 mm
Mass Express
Data Express
Quant Express (optional)
LC express (optional)