SOLATION® ICP-MS
SOLATION® ICP-MS
Inductively coupled plasma mass spectrometry
Model options
- Available models: HM and E
Sample interface
- • Sample Introduction
Peristaltic pump: 4 channels / 12 rollers
Spray chamber: T-controlled, Cyclonic, Quartz
Spray chamber T (°C) (optional): -10 to 10 - Nebulizer: PFA and Quartz
- • Plasma
Torch
Rf generator frequency (MHz): 27
Rf generator power (W): 500-1600
- • Optics
Cones: Cu or Ni
Sampler cone i.d.: 1mm - Skimmer cone i.d.: 0.4mm
Extraction cone i.d.: 3mm
Lens: Einzel
Optics: 90° quad deflector
Mass spectrometer
- • Collision Cell
Collision cell type: Octupole
Kinetic energy discrimination: Yes - No. Gases: 1 (He)
- • Detector
Dwell time (µs): 100
Linear dynamic range: Yes - No. Gases: > 9 orders
- • Mass spectrometers
Mass range (u): 2-260 (212 for E)
Resolution: 0.9 - Scan speed (u/s): 5000
Abundance sensitivity m-1: 1.00E-07
Abundance sensitivity m+1: 1.00E-07
Performance specifications
- • Sensitivity (Kcs/ppb)
7Li: >40
89Y: >300 - 205TI: >100
- • Detection Limits (ppt)
9Be: <1
115ln: <0.5 - 209Bi: <0.5
- • Other Specifications
CeO/Ce (%): <1.5
Ba++/Ba+ (%): <3 - Background cps (@ m/z 220): <1
Stability, 20 min (%): <3
Stability, 2 hours (%): <4
107Ag/109Ag (%): <0.5
Installation requirements
- • Installation, facility requierements
Size: W x D x H (cm): 90 x 66 x 70 cm
Volume (m3): 0.42 - Weight (Kg): 115
- Argon (psi): 100 (99.95% Ar)
- Cell gas (psi): 40 (99.999% He)
- • Environmental requierements
Operating Temperature: 15 to 30 °C (59 – 86 °F). Performance specifications require +/- 2 °C from calibration temperature
Operating Humidity: 10% and 80% RH, non-condensing
Dimensions & weight
- Dimensions:
(WxDxH) 900 x 660 x 700 mm
Weight: 115 kg
Peripheral Equipment & Sample Inlets
- • Autosampler
- Simplifying sample introduction, increasing throughput
- The ASX-560 Autosampler offers:
– Optional dual rinse station
– Optimal chemical compatibility
– Redefined and configurable XYZ movement
– Optimal chemical compatibility Redefined and configurable XYZ movement
- • Sample Racks
– Vials: 90 – Opening: 13.3 mm round – Vial compatibility: 8 mL round bottom or 7 mL flat bottom
– Vials: 60 – Opening: 17.0 mm square – Vial compatibility: 14 mL or 15 mL
– Vials: 21 – Opening: 30.5 mm square – Vial compatibility: 50 mL
- • Rapid Sample Introduction
- More sample throughput, less time
- The ASXPRESS PLUS rapid sample introduction accessory reduces time required for autosampler movement, sample uptake, stabilization and rinse operations, thereby reducing sample run times significantly.
Sample Introduction
1. Autosampler Movement : 5 sec.
2. Sample Uptake: 15 sec.
3. Stabilization: 20 sec.
4. Measurement: 10 sec.
5. Rinse: 30 sec.
Total Time: 80 sec.
1. Autosampler Movement, Sample Uptake, Stabilization and Rinse: 20 sec.
2. Measurement: 10 sec.
Total Time: 30 sec.