SOLATION® ICP-MS

Inductively coupled plasma mass spectrometry
AIS-MassSpec-Puriflash

Model options

  • Available models: HM and E

Sample interface

  • • Sample Introduction
    Peristaltic pump: 4 channels / 12 rollers
    Spray chamber: T-controlled, Cyclonic, Quartz
    Spray chamber T (°C) (optional): -10 to 10
  • Nebulizer: PFA and Quartz

 

  • • Plasma
    Torch
    Rf generator frequency (MHz): 27
    Rf generator power (W): 500-1600

 

  • • Optics
    Cones: Cu or Ni
    Sampler cone i.d.: 1mm
  • Skimmer cone i.d.: 0.4mm
    Extraction cone i.d.: 3mm
    Lens: Einzel
    Optics: 90° quad deflector

Mass spectrometer

  • • Collision Cell
    Collision cell type: Octupole
    Kinetic energy discrimination: Yes
  • No. Gases: 1 (He)

 

  • • Detector
    Dwell time (µs): 100
    Linear dynamic range: Yes
  • No. Gases: > 9 orders

 

  • • Mass spectrometers
    Mass range (u): 2-260 (212 for E)
    Resolution: 0.9
  • Scan speed (u/s): 5000
    Abundance sensitivity m-1: 1.00E-07
    Abundance sensitivity m+1: 1.00E-07

Performance specifications

  • • Sensitivity (Kcs/ppb)
    7Li: >40
    89Y: >300
  • 205TI: >100

 

  • • Detection Limits (ppt)
    9Be: <1
    115ln: <0.5
  • 209Bi: <0.5

 

  • • Other Specifications
    CeO/Ce (%): <1.5
    Ba++/Ba+ (%): <3
  • Background cps (@ m/z 220): <1
    Stability, 20 min (%): <3
    Stability, 2 hours (%): <4
    107Ag/109Ag (%): <0.5

 

Installation requirements

  • • Installation, facility requierements
    Size: W x D x H (cm): 90 x 66 x 70 cm
    Volume (m3): 0.42
  • Weight (Kg): 115
  • Argon (psi): 100 (99.95% Ar)
  • Cell gas (psi): 40 (99.999% He)

 

  • • Environmental requierements
    Operating Temperature: 15 to 30 °C (59 – 86 °F).  Performance specifications require +/- 2 °C from  calibration temperature
    Operating Humidity: 10% and 80% RH, non-condensing

    Dimensions & weight

    Mass spectrometers puriFlash MS purification
    • Dimensions:
      (WxDxH) 900 x 660 x 700 mm
      Weight: 115 kg
    Mass spectrometers puriFlash MS purification

    Peripheral Equipment & Sample Inlets

    ASAP-Probe-fond
    • • Autosampler
    • Simplifying sample introduction, increasing throughput
    • The ASX-560 Autosampler offers:
      –  Optional dual rinse station
      – Optimal chemical compatibility
      – Redefined and configurable XYZ movement
      – Optimal chemical compatibility Redefined and configurable XYZ movement

     

    • • Sample Racks
      – Vials: 90 – Opening: 13.3 mm round – Vial compatibility:  8 mL round bottom or 7 mL flat bottom
      – Vials: 60 – Opening: 17.0 mm square – Vial compatibility:  14 mL or 15 mL
      – Vials: 21 – Opening: 30.5 mm square – Vial compatibility:  50 mL

      ASAP-Probe-fond
      • • Rapid Sample Introduction
      • More sample throughput, less time 
      • The ASXPRESS PLUS rapid sample introduction accessory reduces time required for autosampler movement, sample uptake, stabilization and rinse operations, thereby reducing sample run times significantly.

        Sample Introduction

        1. Autosampler Movement : 5 sec.

        2. Sample Uptake: 15 sec.

        3. Stabilization: 20 sec.

        4. Measurement: 10 sec.

        5. Rinse: 30 sec.

        Total Time: 80 sec.

         

        1. Autosampler Movement, Sample Uptake, Stabilization and Rinse: 20 sec.

        2. Measurement: 10 sec.

        Total Time: 30 sec.